GLOBUS MEDICAL, INC. RESONATE; RESONATE PLATE Back to Search Results
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Insufficient Information (4580)
Event Date 02/20/2024
Event Type  malfunction  
Event Description
It was reported that screws are backing out of a resonate plate post operatively.
 
Manufacturer Narrative
The device was not available for evaluation as it remains in the patient.It was reported that the a screw or screws backed out past the locking mechanisms at the inferior level of a 2 level resonate plate.The original surgery date is not known at this time.It was reported that the back-out was discovered on the 6 month post-op follow up.It was described as the head of the screw slightly protruding from the plate, being barely visible, but appeared to be slightly backed out.The screws were 18mm.According to the surgeon the patient was non-symptomatic and fused.It is unclear whether some damage occurred to the plate or sliders inter-operatively or what the cause of the issue could have been.No pain or adverse effect to the patient was reported.There were no plans for a revision at the time the issue was reported.No determinations can be made for the cause of the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE
Type of Device
RESONATE PLATE
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key18924827
MDR Text Key337916142
Report Number3004142400-2024-00054
Device Sequence Number1
Product Code KWQ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received02/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-