Model Number 0998-00-3013-45
Device Problems Inability to Auto-Fill (1044); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2024
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use, the cs100 intra-aortic balloon pump (iabp) had a circuit leakage alarm and an automatic inflation failure occurred after restarting.The equipment was deactivated and no injuries were caused.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, d9, e1, g3, g6, h2, h3, h4, h6, h10, h11.Due to character restriction block e1 event site telephone is (b)(6).A getinge field service engineer (fse) was dispatched to evaluate the unit.The (fse) reported that the issue was confirmed through the unit logs.The fse replaced the safety disk, and the unit passed all functional and safety tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, CHINESE, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18928611
MDR Text Key337955575
Report Number2249723-2024-01156
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107790
UDI-Public10607567107790
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3013-45
Device Catalogue Number0998-00-3013-45
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2011
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN.
-
-