E1: initial reporter phone: (b)(6).Device evaluated by mfr: synergy xd mr ous 3.50 x 48mm stent delivery system (sds), was returned for analysis.Visual / tactile inspection identified that the stent had been detached from the delivery system and not returned.No issues identified with the hypotube shaft.A visual and tactile examination of the outer and inner lumen and mid-shaft section identified no issues.Microscopic analysis the balloon cones were reviewed, and no issues were noted.Bumper tip showed no signs of distal tip damage.Dimensional testing the device was returned without the stent attached.The stent od (outer diameter) at the time of manufacturing was 0.0437, this is within maximum crimped stent profile measurement.
|
Reportable based on devices analysis completed on 29feb2024.It was reported that crossing difficulties occurred.The 80% stenosed target lesion was located in the severely tortuous and mildly calcified left circumflex artery (lcx).A 3.50 x 48mm synergy xd drug eluting stent was advanced for treatment.However, during the procedure, the device failed to cross the lesion site due to angulation.The procedure was completed with another of the same device.There were no patient complications reported and the patient was stable.However, returned device analysis revealed the stent was detached from the delivery system.
|