Trackwise#: (b)(4).Updated sections: b4, g3, g6, h2, h3, h6, h10.The device was returned to the factory for evaluation on 03/11/2024.An investigation was conducted on 03/14/2024.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.The cord was observed to be intact, with no visual defects observed.An electrical evaluation was performed.The returned cable was connected to a reference power supply and a reference hemopro device.A pre-cautery test was conducted per the instruction for use (ifu) to test the ability of the extension cable to deliver energy.The cable and reference hemopro tool passed the pre-cautery test.It produced visible steam during several activations over a period of 10 minutes and shut off when the toggle was released.Based on the returned condition of the device and the results of the evaluation, the reported failure mode "failure to deliver energy" was not confirmed.The reported device is an oem device, therefore, a lot/serial history review was not applicable.A lot/serial number was not provided and the specific product lot/serial number cannot be identified from a ship history, therefore it is impossible to obtain a certificate of conformance.
|