• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR SUPERA; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR SUPERA; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number UNK SUPERA
Device Problem Stretched (1601)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3: date of event estimated.D4: the udi number is not known as the catalogue and lot number were not provided.D6a: implant date estimated.The device was not returned for analysis.A review of the lot history record and similar incident review of the reported lot could not be conducted because the part and lot number was not provided.It is possible that during stent implantation interaction with the anatomy, other devices and or pre-dilatation was not enough in some locations causing the stent to stretch, elongate after deployment; however, this could not be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.Na.
 
Event Description
It was reported in a general comment that during use of supera stent delivery system (sds), ), it is not known where the stent will land as the stent elongates too much.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA
Type of Device
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18936681
MDR Text Key338084401
Report Number2024168-2024-03512
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
PMA/PMN Number
P120020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK SUPERA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-