SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. CADD ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 21-7022-24
Device Problems Infusion or Flow Problem (2964); Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/13/2024
Event Type  malfunction  
Event Description
It was reported that during use, the line had air bubbles and no liquid was visible.The line was primed with gravity and a chemolock spiros was used and attached at the end of the line during the infusion.Reservoir volume 13.9ml; continuous infusion rate 4ml/hr; total volume on label 184.96ml; total given 174.15ml.Per the reporter, there were no adverse patient effects.
 
Manufacturer Narrative
H3: device not received by manufacturer.No product was returned.The reported complaint could not be confirmed.If the product is returned this complaint will be reopened for further investigation.No lot number was provided; therefore, a history record review could not be conducted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD ADMINISTRATION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18936991
MDR Text Key338312968
Report Number9617604-2024-00240
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586019791
UDI-Public10610586019791
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K895046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number21-7022-24
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/20/2024
Initial Date FDA Received03/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CADD-SOLIS VIP AMBULATORY INFUSION PUMP; CHEMOLOCK SPIROS
-
-