A steris service technician arrived onsite to inspect the 3085 sp surgical table and found that the table's floor locks required repairs.The table was installed in 2001 making it approximately 23 years old and is not under steris service agreement; the user facility is responsible for all maintenance activities.The 3085 sp surgical operator manual states (6-1), "3.3 check floor lock system; have qualified service technician adjust if needed.6x per year" the technician made the necessary repairs to the floor locks, tested the table, confirmed it to be operating according to specification, and returned it to service.The technician counseled user facility personnel on proper preventive maintenance of their 3085 sp surgical table.No additional issues have been reported.
|