• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDLINE EXCELSIOR; SYRINGE, 10ML SALINE IN 10ML

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDLINE EXCELSIOR; SYRINGE, 10ML SALINE IN 10ML Back to Search Results
Model Number EMZE010001H
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/24/2019
Event Type  malfunction  
Manufacturer Narrative
It was reported that the cap of the saline flush syringe was difficult to remove.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.Evaluation of the reported problem/issue was performed using a returned sample and retained samples.Tip cap torque testing was performed on all samples and all syringes were found to be within specification.In an abundance of caution, and in response to an fda 483 issued for cfn 1417592 on 22-jan-2024, this medwatch is being filed.If additional relevant information becomes available a supplemental medwatch will be filed.
 
Event Description
It was reported that the cap of the saline flush syringe was difficult to remove.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SYRINGE, 10ML SALINE IN 10ML
Manufacturer (Section D)
MEDLINE EXCELSIOR
1933 heck drive
neptune NJ 07753 4428
Manufacturer Contact
nigel vilches
three lakes drive
northfield, IL 60093-2753
MDR Report Key18937350
MDR Text Key338241452
Report Number2027791-2024-00038
Device Sequence Number1
Product Code NGT
UDI-Device Identifier00363807100015
UDI-Public00363807100015
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberEMZE010001H
Device Lot Number3133787
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/24/2019
Initial Date FDA Received03/19/2024
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-