Model Number PVF-S
Device Problem Leak/Splash (1354)
Patient Problem Aortic Valve Insufficiency/ Regurgitation (4450)
Event Date 02/19/2024
Event Type  Injury  
Event Description
On (b)(6) 2024, a perceval plus valve pvf-s was attempted to be implanted as follows: patient presented with severe aortic stenosis.Annulus measurement of 21.4, stj of 26.5-27.4 via tee.The usual aortotomy was conducted as well as debridement of the calcium.Sizing was conducted.Clear end of a small sizer fell through, white passed with little resistance.After the small valve was implanted, they were unable to get pressures up above 79mm and seeing a small central leak on tee.As such, it was decided to recess the valve to swap for a sutured valve.As reported, upon opening and perceval removal, there was some additional calcium present on the annulus.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEVAL PLUS SUTURELESS AORTIC HEART VALVE
Type of Device
TISSUE HEART VALVE
Manufacturer (Section D)
CORCYM CANADA CORP.
5005 north fraser way
burnaby, bc V5J 5 M1
CA  V5J 5M1
MDR Report Key18937895
MDR Text Key338068313
Report Number3019892983-2024-01002
Device Sequence Number1
Product Code LWR
UDI-Device Identifier00896208000795
UDI-Public(01)00896208000795(240)PVF-S(17)270113
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 02/19/2024,03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberPVF-S
Device Catalogue NumberPVF-S
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/19/2024
Device Age1 YR
Event Location Hospital
Date Report to Manufacturer02/19/2024
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/19/2024
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-