COVIDIEN MFG DC BOULDER FORCE FX; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number FORCEFX-8C
Device Problems Smoking (1585); Electrical Power Problem (2925); Electrical Shorting (2926); Adverse Event Without Identified Device or Use Problem (2993); Intermittent Energy Output (4025)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2023
Event Type  malfunction  
Event Description
According to the reporter, during use, the unit suddenly lost power, and a burnt smell was smelled.A short circuit was suspected.The department contacted the equipment department, and the engineer went to the site to disassemble the machine and check the fuse.It was found that both fuses were burned out, and a short circuit was suspected.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
According to the reporter, during use, the unit suddenly lost power, and a burnt smell was smelled.A short circuit was suspected.The department contacted the equipment department, and the engineer went to the site to disassemble the machine and check the fuse.It was found that both fuses were burned out, and a short circuit was suspected.There was no patient injury.
 
Manufacturer Narrative
Additional information: b5, e1(first name, last name), e3, g3, h6 this event has been reassessed and found not to be a reportable event and is not associated with a serious injury or potential for serious injury with reoccurrence.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORCE FX
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18938099
MDR Text Key338105489
Report Number1717344-2024-00772
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10884524001562
UDI-Public10884524001562
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K944602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFORCEFX-8C
Device Catalogue NumberFORCEFX-8C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/22/2024
Initial Date FDA Received03/19/2024
Supplement Dates Manufacturer Received03/20/2024
Supplement Dates FDA Received04/16/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-