• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION UNI-VENT, MODEL 731; VENTILATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION UNI-VENT, MODEL 731; VENTILATOR Back to Search Results
Model Number 799-EMBP-05-01
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/29/2024
Event Type  malfunction  
Event Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device displayed "internal comm failure -1173", "internal comm failure -1471", and "internal comm failure -1475" error messages.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the forensic memory log.The device was put through extensive testing without duplicating the report.An internal inspection identified that the cpu to pim board ribbon cable was slightly damaged which may have contributed to the customer report.The cpu to pim ribbon cable was replaced as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNI-VENT, MODEL 731
Type of Device
VENTILATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key18939205
MDR Text Key338075260
Report Number1220908-2024-01087
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00847946024185
UDI-Public00847946024185
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number799-EMBP-05-01
Device Catalogue Number799-EMBP-05-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2024
Was the Report Sent to FDA? No
Date Manufacturer Received02/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-