Model Number 419129
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2024
Event Type  malfunction  
Event Description
An orsiro mission drug-eluting stent system was selected for treatment of a severely calcified lesion (90 percent stenosis degree) in the distal rca.Without predilation the affected device was introduced into the patients body and was unable to pass the lesion.The intervention was finalized without any other treatment.No further information was provided.
 
Manufacturer Narrative
Combination product: yes.Neither the affected device nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.It should be noted that the ifu states to pre-dilate the lesion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO MISSION 3.5/22
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key18944189
MDR Text Key338762873
Report Number1028232-2024-01550
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier07640130441897
UDI-Public(01)07640130441897(17)240522(10)05221674
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P170030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number419129
Device Catalogue NumberSEE MODEL NO.
Device Lot Number05221674
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/15/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial