MAUDE Adverse Event Report: ABBOTT MEDICAL EPIC¿ SUPRA VALVE (AORTIC); HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL EPIC¿ SUPRA VALVE (AORTIC); HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Catalog Number ESP100-21
Device Problem Patient Device Interaction Problem (4001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2024
Event Type  malfunction  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6)2024, a 21mm epic supra valve was chosen for a minimally invasive aortic valve surgery (micsavr).The annular dimension of the native valve was 20.8mm.During procedure, it was noted that the device was tight and did not fall in the annulus and the physician felt resistance felt while parachuting.The device was removed and a new 19mm non- abbott valve was implanted while patient was on bypass.There was no delay in the procedure.The patient was reported stable.
 
Manufacturer Narrative
An event of valve being tight and resistance felt while parachuting was reported.Information from field indicated that the annular dimension of the native valve was 20.8 mm and the 21 mm valve was removed and downsized to 19 mm non-abbott valve.A returned device assessment could not be performed as the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Event Description
It was reported that on (b)(6) 2024, a 21mm epic supra valve was chosen for a minimally invasive aortic valve surgery (micsavr).The annular dimension of the native valve was 20.8mm.During procedure, it was noted that the device was tight and did not fall in the annulus and the physician felt resistance felt while parachuting.The device was removed and a new 19mm non- abbott valve was implanted while patient was on bypass.There was no delay in the procedure.The patient was reported stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC¿ SUPRA VALVE (AORTIC)
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18944650
MDR Text Key338434876
Report Number2135147-2024-01257
Device Sequence Number1
Product Code LWR
UDI-Device Identifier05414734027465
UDI-Public(01)05414734027465(17)270828(10)10099694
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P040021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberESP100-21
Device Lot Number10099694
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/20/2024
Supplement Dates Manufacturer Received05/02/2024
Supplement Dates FDA Received05/07/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/29/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age21 YR
Patient SexFemale
-
-