Model Number CDHFA500Q
Device Problems Failure to Interrogate (1332); Wireless Communication Problem (3283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  malfunction  
Event Description
It was reported that the patient's implantable cardioverter defibrillator exhibited loss of bluetooth telemetry.No intervention was performed.There were no patient consequences.
 
Event Description
New information received notes that the patient's implantable cardioverter defibrillator was operating in bluetooth telemetry lockout.The issue was resolved via reinterrogation in clinic.There were no patient consequences.
 
Manufacturer Narrative
The reported event of bluetooth telemetry loss was verified.Information received indicates that the patient's implantable cardioverter defibrillator had gone into telemetry lockout, which is indicative of normal device function.No anomalies were detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18946779
MDR Text Key338193517
Report Number2017865-2024-35497
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 05/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Model NumberCDHFA500Q
Device Lot NumberS000078341
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/20/2024
Supplement Dates Manufacturer Received03/27/2024
05/22/2024
Supplement Dates FDA Received04/16/2024
05/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/11/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA.; QUARTET.; TENDRIL.
Patient SexMale
-
-