• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 32" (81 CM) APPX 6.3 ML, 15 DROP ADMIN SET W/0.2 MICRON FILTER, ROTATING LUER W/; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 32" (81 CM) APPX 6.3 ML, 15 DROP ADMIN SET W/0.2 MICRON FILTER, ROTATING LUER W/; STOPCOCK, I.V. SET Back to Search Results
Catalog Number B30183
Device Problem Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2024
Event Type  malfunction  
Manufacturer Narrative
One (1) used sample list #b30183 connected with a used, equashield, and used, bag spike adaptor; lot #unknown and one used.A 0.9% sodium chloride 500ml bag was returned for evaluation.As received the inlet filter appeared to be wetted out.No additional damage or anomalies were observed.The set was primed and purged as per procedure and a leak coming from the inlet filter vent was confirmed.However, no air beyond the outlet filter vent to the patient line was confirmed.Complaints of product incorporates/allows air passage cannot be confirmed.However.The probable cause of the leaks coming from the inlet filter is typically a temporary or complete loss of hydrophobic properties of the filter vent material due to an infusate interaction during use.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.
 
Event Description
It was reported that a 32" (81 cm) appx 6.3 ml, 15 drop admin set w/0.2 micron filter, rotating luer w/filter cap, bag hanger allowed air to enter the tubing, trapping fluid above in the bag and drip chamber during patient use.Then air in the tubing down to the filter and there is fluid below the filter.They are unable to back prime to clear the line of air and run the medicine.The device is contaminated with chemotherapy.There was a patient involved, no patient harm and there was a delay in therapy as the drug had to be remade to complete the therapy.No additional information is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32" (81 CM) APPX 6.3 ML, 15 DROP ADMIN SET W/0.2 MICRON FILTER, ROTATING LUER W/
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18947294
MDR Text Key338328699
Report Number9617594-2024-00303
Device Sequence Number1
Product Code FMG
UDI-Device Identifier00887709055445
UDI-Public(01)00887709055445(17)281101(10)13828709
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K964435
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB30183
Device Lot Number13828709
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/19/2024
Initial Date FDA Received03/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNSPECIFIED CHEMO, UNK MFR
-
-