• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102812-81A
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2024
Event Type  malfunction  
Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Event Description
On february 22, 2024, senseonics was made aware of an incident where user received an early sensor replacement alert resulting in an early sensor removal.
 
Manufacturer Narrative
The sensor replacement alert was first presented to the user on (b)(6) 2024, day 141 post insertion.The sensor was tested in-house, and the review of investigation analysis revealed a loss of chemical performance.The system correctly disabled the sensor when it detected the performance failure, and the system's self-test functions were working normally.The root cause of the performance failure was due to the sensor's hydrogel oxidation.As part of resolution, a return material authorization was issued for sensor replacement.No further resolution was necessary for this complaint.B4.Date of this report 14 may 2024.D9 device available for evaluation? yes, 12 april 2024.G3.Date received by the manufacturer? 10 may 2024.H3.Device evaluated by manufacturer? yes.H6.Type of investigation updated to 10.H6.Investigation findings updated to 3231.H6.Investigation conclusions updated to 4307.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key18947366
MDR Text Key338208225
Report Number3009862700-2024-00090
Device Sequence Number1
Product Code QHJ
UDI-Device Identifier00817491022974
UDI-Public817491022974
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/03/2023
Device Model Number102812-81A
Device Catalogue NumberFG-7202-00-301
Device Lot NumberWP09134
Was Device Available for Evaluation? No
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
-
-