Model Number C40+
Device Problem Circuit Failure (1089)
Patient Problem Failure of Implant (1924)
Event Date 02/23/2024
Event Type  malfunction  
Manufacturer Narrative
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event Description
The user's hearing performance with the device was affected.The user was re-implanted on (b)(6) 2024.
 
Manufacturer Narrative
Conclusion: device investigation confirmed that the stimulator electronics is not working according to specifications.Based on the manufacturer_s experience with this kind of devices, it can be assumed that the device has failed due to loss of hermeticity at the housing braze joint.The investigation results appear to match the problems mentioned in the recipient report.This is a final report.
 
Event Description
The user's hearing performance with the device was affected.The user was re-implanted on (b)(6) 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key18947616
MDR Text Key338207646
Report Number9710014-2024-00259
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberC40+
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/14/2024
Initial Date FDA Received03/21/2024
Supplement Dates Manufacturer Received03/14/2024
Supplement Dates FDA Received05/08/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age28 YR
Patient SexMale
-
-