MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS VASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS VASCULAR STENT GRAFT Back to Search Results
Catalog Number FVL14100
Device Problems Positioning Failure (1158); Misfire (2532)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2024
Event Type  malfunction  
Event Description
It was reported that during a stent graft procedure through common iliac artery, the device allegedly failed to deploy.Reportedly, the defected stent was removed.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the fluency plus endovascular stent graft that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the stent delivery system was returned for evaluation and the stent graft was found partially deployed which leads to confirmed results for partial deployment.Based on available information and evaluation of the returned sample, the investigation is closed with confirmed results for partial deployment.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.The instructions for use states "if excessive force is felt during stent graft deployment, do not force the delivery system.Remove the delivery system and replace with a new unit".Regarding preparation of the device the instructions for use states: "prior to loading the delivery system over a guide wire, both ports must be flushed with sterile saline to eliminate any air bubbles that may be trapped in the inner catheter lumen and/or the stent graft lumen.Flushing these lumens will also facilitate stent graft deployment.A super stiff guide wire (0.035 in.) is advanced from a femoral artery puncture site.Use an introducer sheath for the implant procedure".The packaging pictograms indicate an introducer size of 10f and a 0.035" guidewire.H10: d4 (expiration date: 09/2025) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS VASCULAR STENT GRAFT
Type of Device
VASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18947828
MDR Text Key338194365
Report Number9681442-2024-00056
Device Sequence Number1
Product Code PFV
UDI-Device Identifier00801741145360
UDI-Public(01)00801741145360
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P130029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFVL14100
Device Lot NumberANGW3185
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/13/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-