BIO-RAD MEDICAL DIAGNOSTICS GMBH REAGENT RED BLOOD CELLS IH-PANEL 11; IH-PANEL 11; 11X4 ML Back to Search Results
Catalog Number 814070100
Device Problem False Positive Result (1227)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/24/2024
Event Type  malfunction  
Manufacturer Narrative
This is our initial report on this incident.
 
Event Description
The customer reported a false positive reaction of cell no.8 (donor no.103493) of ih-panel 11 with anti-n.According to the antigen table cell no.8 was n antigen negative, but mmss.The customer stated that they received a positive reaction with cell no.8 of ih-panel 11 and seraclone anti-n.The customer intends to return a product sample for investigational tesing but hasn't done so yet.Therefore, our quality control laboratory tested their retention sample of the supposedly defective lot.Cell no.8 of ih-panel 11 was tested with seraclone anti-m and seraclone anti-n according to the instruction for use (tube test, incubation for 30 minutes at room temperature, reading of the results; no centrifugation).Cell no.8 of ih-panel 11 showed a correctly negative result with seraclone anti-n, while the reaction with seraclone anti-m was correctly positive.The positive and negative controls also showed a correct result.We did not observe any false positive reaction.Based on the current tests of our quality control laboratory, a mistyping of cell no.8 of ih-panell 11 (donor no.103493) can be excluded.The donor 103493 has already been in use for the paned reagent red blood cells for four years.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.
 
Manufacturer Narrative
This is our final report on this incident.
 
Event Description
The customer reported a false positive reaction of cell no.8 (donor no.(b)(6)) of ih-panel 11 with anti-n.According to the antigen table cell no.8 was n antigen negative, but mmss.The customer stated that they received a positive reaction with cell no.8 of ih-panel 11 and seraclone anti-n.First, our quality control laboratory tested their retention sample of the supposedly defective lot.Cell no.8 of ih-panel 11 was tested with seraclone anti-m and seraclone anti-n according to the instruction for use (tube test, incubation for 30 minutes at room temperature, reading of the results; no centrifugation).Cell no.8 of ih-panel 11 showed a correctly negative result with seraclone anti-n, while the reaction with seraclone anti-m was correctly positive.The positive and negative controls also showed a correct result.We did not observe any false positive reaction.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.The customer sent in a product sample for investigational testing.When we received the material returned by the customer, testin at our quality control laboratory confirmed that the customer vial of ih-cell no.8 reacted correctly negative with seraclone anti-n.This was also confirmed with other m+n- red cells.But when the ih-cell no.8 was used with ih-liss and seraclone anti-n in the tube test as the customer did, a false positive reaction occurred.However, this use was not in accordance with the ifu, because ih-panel 11 and ih-liss are for use in the ih system, while the intended use of seraclone anti-n is the tube test.A false positive result only occurred in this use against the ifu.Otherwise, cell no.8 of ih-panel 11 (donor no.(b)(6)) showed correct results.In addition, the potency and specificity of the customer's seraclone anti-n was tested and confirmed in comparison to the retention sample.Based on the investigation the complaint was classified as unjustified.Both the complaint sample and the retention sample reacted as expected.Contrary to the ifu, the customer had prepared the ih cells with ih-liss (also for the ih system and not the tube method) for use in the tube method and the use of seraclone anti-n and obtained false positive results.A false positive result only occurred in this use not according to the ifu.The combination of red blood cells suspended in ih-liss and seraclone anti-n which did not comply with the ifu led to the false positive reactions.Otherwise, cell no.8 of ih-panel 11 (donor no.(b)(6)) showed correct results.An incorrect determination of the cell no.8 of ih-panel 11 (donor no.(b)(6)) regarding the mns system can safely be ruled out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAGENT RED BLOOD CELLS IH-PANEL 11
Type of Device
IH-PANEL 11; 11X4 ML
Manufacturer (Section D)
BIO-RAD MEDICAL DIAGNOSTICS GMBH
industriestrasse 1
dreieich, hessen 63303
GM  63303
Manufacturer Contact
martina benkert
industriestrasse 1
dreieich, 63303
GM   63303
MDR Report Key18948863
MDR Text Key339211652
Report Number9610824-2024-00005
Device Sequence Number1
Product Code KSZ
UDI-Device Identifier07611969952298
UDI-Public(01)07611969952298(17)240408(10)9406011
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
125208
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/08/2024
Device Catalogue Number814070100
Device Lot Number9406011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/24/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-