• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRO MED INSTRUMENTS GMBH DORO® QR3 SKULL CLAMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRO MED INSTRUMENTS GMBH DORO® QR3 SKULL CLAMP Back to Search Results
Model Number 1001.001
Device Problem Device Slipped (1584)
Patient Problem Laceration(s) (1946)
Event Date 02/22/2024
Event Type  Injury  
Manufacturer Narrative
The product was sent in for inspection in combination with a third-party torque screw.A clamping force inspection is not possible with the third-party component.Any risks arising from the use of a third-party component, in particular the torque screw, cannot be excluded nor assessed.The permissible components that can be combined with the product are specified in the instructions for use of the product.The second deviation, the slight play in the arch holder cannot have contributed to a slippage and has therefore no casual link to described incident.The maintenance interval was exceeded by the customer by 5 months.Any deviations detected could not have gone undetected during routine inspection and maintenance.The maintenance interval of one year is specified as mandatory in the product's instructions for use.No causal link can be established between the skull clamp sent in and the slippage described.In our experience, the pinning technique is decisive for the stability of the fixation of the patient's skull.It cannot be excluded that the pinning technique in this case was not optimal, as described in the instruction manual: "adjust the skull clamp to the width of the patient's head in the manner that the two skull pins in the rocker arm are equidistant from the centerline of the head and the single skull pin at the extension assembly is in line with this centerline.".
 
Event Description
Customer informed us on february 22 that one of our products was involved in a procedure (suboccipital craniotomy) in which the patient sustained a laceration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® QR3 SKULL CLAMP
Type of Device
DORO® QR3 SKULL CLAMP
Manufacturer (Section D)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM  79111
Manufacturer (Section G)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM   79111
Manufacturer Contact
nicholas preissler
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM   79111
MDR Report Key18950458
MDR Text Key338252765
Report Number3003923584-2024-00005
Device Sequence Number1
Product Code HBL
UDI-Device Identifier04250435506196
UDI-Public04250435506196
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191740
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number1001.001
Device Catalogue Number1001.001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2024
Initial Date Manufacturer Received 02/22/2024
Initial Date FDA Received03/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-