• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS® SARS-COV-2 & INFLUENZA A/B; COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS® SARS-COV-2 & INFLUENZA A/B; COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES Back to Search Results
Catalog Number 09211101190
Device Problem False Positive Result (1227)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2024
Event Type  malfunction  
Manufacturer Narrative
The observed discrepancy is most likely due to the sample being a weak positive for the flu a target that is at/near the limit of detection (lod) of the assay.Low viral load specimens that are near the assay lod may not generate consistent results upon repeat testing according to expected statistical variances in detection.A result discrepancy may be expected between assays due to potential differences in the tests¿ limits of detection (lod) and intended use.
 
Event Description
In light of the covid-19 pandemic and the subsequent emergency use authorizations (euas) for sars-cov-2 diagnostic tests, the agency has requested heightened reporting beyond the reasonably suggests requirements of 803 to include allegations of false positive or false negative results independent of harm or malfunction or off-label use.Pursuant to the agency¿s instruction, we hereby submit this mdr.A customer alleged discrepant results for two patients while using the cobas® sars-cov-2 & influenza a/b nucleic acid test on the cobas® liat® system.The alleged samples initially generated a negative result for all targets using the cobas® influenza a/b & rsv assay.The same samples were retested on the same cobas liat analyzer using the cobas® sars-cov-2 & influenza a/b assay which yielded a positive result for influenza a.Sample 2 was repeated a total of 3 times on the same liat yielding the same positive result for influenza a.The repeat results were released.No harm was alleged.An investigation was conducted to evaluate the customer issue.Per fda¿s eua guidance, 2 mdrs will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS® SARS-COV-2 & INFLUENZA A/B
Type of Device
COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES
Manufacturer (Section D)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
branchburg NJ 08876
Manufacturer (Section G)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
branchburg NJ 08876
Manufacturer Contact
timothy blair
1080 us hwy 202 s
branchburg, NJ 08876
MDR Report Key18950893
MDR Text Key338558123
Report Number2243471-2024-00850
Device Sequence Number1
Product Code QLT
Combination Product (y/n)Y
Reporter Country CodeUK
PMA/PMN Number
EUA201779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number09211101190
Device Lot Number21114Z
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/27/2024
Initial Date FDA Received03/21/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-