MEDTRONIC NAVIGATION, INC FUSION® ENT NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT
|
Back to Search Results |
|
Model Number 9733560 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/13/2024 |
Event Type
malfunction
|
Event Description
|
Medtronic received information regarding a navigation device being used for a functional endoscopic sinus surgery (fess).It was reported that the "emitter faulted" was displayed.The site aborted navigation.There was no reported impact to patient outcome and a reported delay of less than one hour.
|
|
Manufacturer Narrative
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: e mitter 9735521 axiem 3 side mount h3) no parts have been returned for analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
H2) continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: (b)(4), serial/lot #: (b)(6).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
H3/h6 - a manufacturer representative went to the site to service the system.The emitter was replaced.Evaluation codes b01, c02, d02 apply.The returned emitter (b)(6) lot# 603005825 was returned for evaluation.Analysis found no fault with the component.Evaluation codes b01, c19, d14 apply.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|