The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
A "replace sensor" message was reported with the abbott diabetes care (adc) device and the customer was unable to obtain readings.The customer experienced weakness and feeling unwell, and was unable to self-treat, requiring non-healthcare professional administration of compresses on the forehead and sweet tea.It was additionally indicated that an ambulance was called, but no further details were provided.There was no report of death or permanent impairment associated with this event.
|
Sensor (b)(6) has been returned and investigated.The sensor plug is fully seated and no issues were observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Inspected the plug assembly and cracked sensor neck was observed.The cause of the cracked sensor neck is likely attributed to shipment of the used sensor back to adc for investigation.Sensor state 5 is an indication of normal sensor termination and full wear by the user, therefore the cracked sensor neck observed during investigation occurred post-wear.In addition, the passing of all tests during the investigation indicates that the cracked sensor neck did not impact the sensor's ability to still generate an accurate glucose reading.Current was applied to the sensor to perform sim vivo testing (simulation of the electrical signal produced by the sensor tail) and poise voltage testing.All results were within specification.Therefore issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
A "replace sensor" message was reported with the abbott diabetes care (adc) device and the customer was unable to obtain readings.The customer experienced weakness and feeling unwell, and was unable to self-treat, requiring non-healthcare professional administration of compresses on the forehead and sweet tea.It was additionally indicated that an ambulance was called, but no further details were provided.There was no report of death or permanent impairment associated with this event.
|