SMITHS MEDICAL ASD, INC. CONTINENTAL EUROPEAN 230 VOLT LEVEL 1 HOTLINE; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Model Number HL-90 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/23/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
E1 additional phone - (b)(6).One device was received in used condition.Per visual inspection, no cracks, or damages visible.Per functional testing, the complaint could not be confirmed.Unit works as expected.The water level sensor works normally.No other problem found.Electrical and safety test passed.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.The device passed all functional and delivery tests.No further actions were taken.
|
|
Event Description
|
It was reported that the device has a faulty level sensor.There was no patient involvement, and no harm/adverse event reported.
|
|