It was reported that in order to maintain the patient's body temperature during the operation, the convective warming machine was initially kept at 36°c.However, due to subsequent extensive flushing, the patient's body temperature dropped, so it was adjusted to 40°c.The device was not connected as required (but a cloth was wrapped around the hot air outlet).The sensor at the air outlet fell off during use, causing a real-time temperature control failure and the device remained continuously heating.Due to the long duration of the operation, the patient suffered burns over a large area from the back of the right shoulder to the right upper arm, and there were blisters in some areas.Later, engineers from the hospital put the equipment into test, it was found that if the temperature probe is detached and the temperature is set to 44°c (during test), the equipment will continue to heat, but the temperature will remain at about 53°c, and this could cause burns to the patient.
|