It was reported that a pre-filled syringes were found to have leaked within their packaging.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.A sample was returned for evaluation and it was confirmed to have a cracked barrel and a tear in the overwrap packaging.Retention samples were evaluated with no similar problem/issue identified.It is likely that the cracked syringe was caused by low temperatures experienced by the syringe during or after shipment.In an abundance of caution, and in response to an fda 483 issued for cfn 1417592 on 22-jan-2024, this medwatch is being filed.If additional relevant information becomes available a supplemental medwatch will be filed.
|