E1: initial reporter address: (b)(6).H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed which revealed a leak at the spike port bonding area.The reported condition was verified.The cause was not determined; however, a likely cause was due to inadequate, or lack of cyclohexanone being applied to the spike port cap tube when it was inserted to the spike port during the manufacturing process causing an incorrect bonding.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|