Model Number 72081-01
Device Problem High Readings (2459)
Patient Problems Hypoglycemia (1912); Dizziness (2194); Loss of consciousness (2418)
Event Date 02/18/2024
Event Type  Injury  
Event Description
A high readings issue with the adc device was reported.The customer reported unspecified "too high" sensor readings, and lost consciousness.The customer was provided unspecified treatment by a third-party, and additionally taken to the emergency room and admitted to hospital as inpatient.The customer was treated with 1000 ml e153 (electrolyte solution), 250 ml 5% glucose solution, and 12 g 20% glucose for a diagnosis of hypoglycemia.A laboratory blood glucose result of 4.1 mmol/l was reported, taken at unspecified time in relation to treatment.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time, product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kit were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 3
Type of Device
CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE LTD
range road
witney 12345 -700
UK  12345-7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key18956584
MDR Text Key338302168
Report Number2954323-2024-09599
Device Sequence Number1
Product Code QLG
UDI-Device Identifier35021791002659
UDI-Public(01)35021791002659(17)2024-10-31(10)T60001475(91)72114-01
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K213996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number72081-01
Was Device Available for Evaluation? No
Date Manufacturer Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/23/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-