Device evaluation of monitor has been completed.The reported problem (damaged monitor case) was confirmed due to the monitor's electrode belt receptacle being broken free from the monitor enclosure, damaging wires within the connector and causing fault.After incoming evaluation, contamination was found throughout the monitor¿s ca board.The root cause for the damaged belt receptacle was physical abuse.The device is designed to meet iec 60601-1 mechanical requirements.Lifevest patient instructions for use remind and warn patients not to expose the lifevest electronic components to liquids.There was no adverse event that resulted from the damaged monitor.
|