• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRELINK; DATA MANAGEMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRELINK; DATA MANAGEMENT SYSTEM Back to Search Results
Model Number 71892-01
Device Problem Application Program Problem (2880)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/15/2023
Event Type  malfunction  
Manufacturer Narrative
This complaint was investigated.Adc has identified a software defect for the freestyle librelink application for ios, version 2.10.0, in which the application upgrade was unsuccessful.This resulted in a period where users may not have received glucose results or may not have been alerted to low or high glucose alarms.Based on the investigation, the complaint is confirmed and no further investigation activities for this individual complaint are required.This issue was addressed in the field by adc fa1029-2023.The device model number populated in section d4 is for the freestyle librelink ios application, on market in the uk.This is same/similar to us freestyle librelink/freestyle libre 2 ios application part number 71733-01/71926-01.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
An unexpected application error issue was reported with the adc application version 2.10.0, in use with iphone 7 with ios operating system version 15.4.1.Adc has identified that following the release of the freestyle librelink (fsll) application for ios, v2.10.0, on 12-jul-2023 (11am) in the united kingdom (uk), some users have experienced a situation where the application upgrade is unsuccessful, and as a result, they receive a white screen in the application user interface (ui).This issue does not affect users in the united states.This will result in a period where glucose readings and alarms will not be received, and historical glucose readings will not be accessible, as the previous fsll application, v2.8.1, was no longer available to users on the apple app store.This issue was addressed in the field by adc fa1029-2023 and was resolved in the field by the release of updated fsll ios application, made available in the apple app store in the uk on 17-jul-2023.There was no report of death, serious injury, or mistreatment associated with this event.This report is being submitted as part of retrospective reporting deemed necessary related to fa1029-2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRELINK
Type of Device
DATA MANAGEMENT SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key18960559
MDR Text Key338429015
Report Number2954323-2024-09788
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Notification
Type of Report Initial
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number71892-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/15/2023
Initial Date FDA Received03/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberQR908720
Patient Sequence Number1
-
-