THORATEC CORPORATION HEARTMATE 3 SYSTEM CONTROLLER; VENTRICULAR (ASSIST) DEVICE Back to Search Results
Model Number 106531US
Device Problems Disconnection (1171); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  malfunction  
Event Description
It was reported that during a patient visit the system controller was noted to have a driveline power fault, as well as dust/debris in the modular cable connection.The modular cable and controller were exchanged and no further alarms were noted.Log files were submitted for review and contained the onset of driveline power a faults on (b)(6)2024.There were also 2 separate sets of alarms associated with driveline disconnects on (b)(6)2024.It was noted that following the disconnects the driveline fault appeared to have persisted.The driveline disconnects were reportedly not related to troubleshooting, and it was unknown if the modular cable had any trauma, though no visible damage was noted.Related manufacturer's reference number for the pump: 2916596-2024-01494 related manufacturer's reference number for the modular cable: 2916596-2024-01496.
 
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 SYSTEM CONTROLLER
Type of Device
VENTRICULAR (ASSIST) DEVICE
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18963528
MDR Text Key339028417
Report Number2916596-2024-01495
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024013235
UDI-Public00813024013235
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number106531US
Device Lot Number8322711
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/28/2024
Initial Date FDA Received03/22/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age32 YR
Patient SexMale
Patient Weight72 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-