This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.The device was returned to olympus for inspection, and the reported malfunctions were confirmed.A definitive root cause was not identified; however, based on the results of the investigation, the probable cause of the malfunction would likely be the kinking of the image sensor cable, leading to the disconnection and short-circuiting of the output signal cable, as indicated in the investigation report.Additionally, it was confirmed that scratches and chips were present on the glue portion of bending section cover (a-rubber).Therefore, we presume that the kinking of the image sensor cable occurred when the trocar was inserted at a tilt, and significant external stress, such as excessive twisting and bending, was applied to the cable.The event can be prevented by following the instructions for use which state: the instruction manual operation manual "chapter 3 preparation and inspection, 3.8 inspection of the endoscopic system¿.Olympus will continue to monitor field performance for this device.
|