Model Number 71992-01 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problems
Dehydration (1807); Headache (1880); Hyperglycemia (1905); Vomiting (2144); Loss of consciousness (2418); Convulsion/Seizure (4406)
|
Event Date 03/03/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An inaccurate readings issue was reported with the use of the abbott diabetes care (adc) device.The customer reported receiving unspecified inaccurate sensor scan readings and experienced symptoms described as "vomiting, migraine, high blood pressure, extreme dehydration", a seizures, and a loss of consciousness.The customer was unable to self-treat and required an insulin injection(dose/type unspecified) from both a non-healthcare professional and a healthcare professional for the diagnosis hyperglycemia.No further treatment was reported.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect deficiency.If a product defect deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit risk ratio.At this time product has not yet been returned and a valid serial number has not been provided.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Clinical data was reviewed and confirmed that libre sensors continue to be safe, effective, and perform as intended in the field.Stability data for libre sensors was reviewed and showed no anomalies or non-conformances that could have lead to the complaint.A tripped trend review was conducted for the reported complaint and fs libre sensors, no trends were identified that would indicate any product related issues.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An inaccurate readings issue was reported with the use of the abbott diabetes care (adc) device.The customer reported receiving unspecified inaccurate sensor scan readings and experienced symptoms described as "vomiting, migraine, high blood pressure, extreme dehydration", a seizures, and a loss of consciousness.The customer was unable to self-treat and required an insulin injection (dose/type unspecified) from both a non-healthcare professional and a healthcare professional for the diagnosis hyperglycemia.No further treatment was reported.There was no report of death or permanent impairment associated with this event.
|
|