Catalog Number 5803002
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  malfunction  
Manufacturer Narrative
Investigation is in process.A follow-up report will be provided.
 
Event Description
The customer reported that after completing an apheresis platelet procedure on a trima device, the operator noticed that the collected platelet product looked red.The operator tested the product and found that the free hemoglobin content was high.There was not a transfusion recipient or patient involved at the time of the unit processing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.11.Investigation: a photograph was submitted in lieu of the disposables set to aid in the investigation.The photograph shows used primary and secondary platelet bags, and associated tubing, from a trima accel disposable set.The photograph confirms the presence of a red tinge.No kinks or misassemblies are visible.A disposable complaint history search was performed for this lot and found zero reports for similar issues on this lot.Investigation is in process and a follow-up report will be provided.
 
Event Description
The customer reported that after completing an apheresis platelet procedure on a trima device, the operator noticed that the collected platelet product looked red.The operator tested the product and found that the free hemoglobin content was high.There was not a transfusion recipient or patient involved at the time of the unit processing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL PLT, PLASMA, ISBT
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key18965272
MDR Text Key338434674
Report Number1722028-2024-00098
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583803008
UDI-Public05020583803008
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5803002
Device Lot Number2310134151
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/27/2024
Initial Date FDA Received03/23/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-