H11: additional manufacturer narrative: updated: b4, d9, g3, g6, h2, h3, h6 h3: evaluation summary: complaint was of unknown reasons was unable to be confirmed.X-ray demonstrated frame was expanded, deformed and pushed inward around all three commissures.Host tissue was heavy at the frame inflow aspect and minimal at the outflow aspect.As received, all three leaflets had cuts of approximately 10mm x 12mm on leaflet 1, 3mm on leaflet 2, and 12mm x 14mm on leaflet 3.The cuts had a smooth and straight edge; cut leaflet were not returned.Mechanical damage marks were also observed on the outflow aspect of leaflet 2.Damages appeared serrated and did not penetrate leaflets.Sewing ring was cut around the valve and exposed the wireform and metal band on the outflow aspect.Cut off sewing ring fragment was not returned with valve.Wireform was also exposed around commissures 1 and 3.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
|
Through implant patient registry it was learned that a 21mm 8300ab valve in the aortic position was explanted after an implant duration of 4 years, 9 months due to unknown reason.The explanted valve was replaced with a 21mm 11500a valve.Per product evaluation, host tissue was heavy at the frame inflow aspect and minimal at the outflow aspect.Mechanical damage marks were also observed on the outflow aspect of leaflet 2.
|