• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH CARDIOHELP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH CARDIOHELP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number CARDIOHELP-I
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  malfunction  
Manufacturer Narrative
A follow up will submitted when additional information become available.
 
Event Description
It was reported that the cardiohelp showed the alarm ¿pump disposable error¿ during treatment.No harm to any person has been reported.Complaint id: (b)(4).
 
Event Description
Complaint id: (b)(4).
 
Manufacturer Narrative
A getinge field service technician (fst) was sent for investigation the failure could not be replicated and no parts were replaced.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.A follow up will submitted when additonal information become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOHELP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key18967010
MDR Text Key338428480
Report Number8010762-2024-00159
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCARDIOHELP-I
Device Catalogue Number701048012
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/19/2024
Initial Date FDA Received03/25/2024
Supplement Dates Manufacturer Received04/16/2024
Supplement Dates FDA Received04/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-