SMITH & NEPHEW, INC. GII PATELLAR REAMER SHAFT; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Catalog Number 71440324 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/04/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that, during a tka procedure, the silver ring of a gii patellar reamer shaft disassociated during bcs femoral reaming.It is unknown how procedure was resumed.The procedure was completed, without any delay.Patient was not harmed as consequence of this problem.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Manufacturer Narrative
|
H3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.H10- additional information h8- usage of device h11- corrected data b5- describe event or problem d8- device available for evaluation.
|
|
Event Description
|
It was reported that, during a tka procedure, the silver ring of a gii patellar reamer shaft disassociated during bcs femoral reaming.It is unknown how procedure was resumed, but it was completed without any delay.Patient was not harmed as consequence of this problem.
|
|