• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2024
Event Type  malfunction  
Event Description
It was reported that the patient presented to the hospital for a scheduled implant procedure.During the procedure, the left ventricular (lv) lead was found difficult to be separated from the guidewire.Upon removal of the guidewire from the lv lead, the distal tip of the guidewire was found broken.The physician elected to remove and replace the lv lead as a precautionary measure on (b)(6) 2024.The patient was in stable condition throughout.
 
Manufacturer Narrative
The reported event difficult to remove stylet/guidewire was confirmed.As received, a complete lead was returned in one piece.Visual inspection and x-ray examination to the distal region and the entire lead found no anomalies, but the inner coil was bunched up at the connector region, consistent with procedure damage.Electrical testing found no conductor fractures or internal shorts.The cause of the reported event was due to inner coil damage due to procedural damage.No other anomalies were seen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18976374
MDR Text Key338593282
Report Number2017865-2024-35860
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberA000152642
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/06/2024
Initial Date FDA Received03/25/2024
Supplement Dates Manufacturer Received04/01/2024
Supplement Dates FDA Received04/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
Patient Weight81 KG
-
-