An alarm issue was reported with the adc device.The customer experienced a signal loss and was unable to obtain readings and receive glucose alarms.As a result, customer was not alerted of changes in glucose level and experienced cold sweat, dizziness and was unable to self-treat.The customer had contact with a healthcare professional (hcp) who administered unspecified treatment for a hyperglycemia diagnosis.There was no report of death or permanent impairment associated with this event.
|
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
An alarm issue was reported with the adc device in use with zte a73 5g phone with an android 13 operating system with app version 3.5.1.10363.The customer experienced a signal loss and was unable to obtain readings and receive glucose alarms.As a result, customer was not alerted of changes in glucose level and experienced cold sweat, dizziness and was unable to self-treat.The customer had contact with a healthcare professional (hcp) who administered unspecified treatment for a hyperglycemia diagnosis.There was no report of death or permanent impairment associated with this event.
|
Sensor (b)(6) has been returned and investigated.Visual inspection performed on the returned sensor and de-cased sensor was observed.An internal visual inspection was performed on the sensor¿s pcba (printed circuit board assembly); no issues were observed.The returned battery voltage was measured to be within specification range.The returned sensor was re-programmed to storage state and activated for smu (source measurement unit) testing.Performed an smu (source measurement unit) test to ensure the sensor's electronics were functioning correctly, and the returned unit did not have any glucose reading issues.The returned sensor was re-programmed to activate with known good reader.After waiting for few minutes, the known good reader was connected to masterm and command was sent to isf (interstitial fluid) and first 5 ble(bluetooth) packets were received.Therefore, this issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|