C.R. BARD INC. (COVINGTON) -1018233 LEG BAG Back to Search Results
Catalog Number 150832
Device Problem Restricted Flow rate (1248)
Patient Problem Discomfort (2330)
Event Date 03/05/2024
Event Type  malfunction  
Event Description
It was reported that the patient called about issue they had been experiencing with the catheter.The patient stated that the urine was not draining through the tube and was creating negative pressure which was causing discomfort at the tip of penis.It was stated that the patient was concerned about infection with this.The representative informed to the patient to try running water through the leg bag first, because it was like a vacuum and causing the backflow.No medical intervention was reported.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.The potential root cause for this failure could be due to "poor interfaces with connections".The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "directions for use: 1.Separate notches within circles.Pull straps through holes and around leg.2.Position bag on leg with flutter valve at top.3.Attach catheter or extension tubing to top inlet.When wearing bag below knee, attach bard® extension tubing (catalog no.150615 or 4a4194).When using extension tubing, make sure to connect tube at least to the 3rd taper of the connector.4.To empty dispoz-a-bag®, push green lever on flip-flo¿ valve out and down.Important: be sure to reclose flip-flo¿ valve after emptying bag.5.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable, local, state and federal laws and regulations." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEG BAG
Type of Device
LEG BAG
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18977836
MDR Text Key338587284
Report Number1018233-2024-01576
Device Sequence Number1
Product Code FAQ
UDI-Device Identifier00801741050886
UDI-Public(01)00801741050886
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number150832
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-