• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97715
Device Problem High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  Injury  
Event Description
Information was received from a patient via manufacturer representative who was implanted with an implantable neurostimulator (ins).It was reported that there were high impedances, 40000.They tried programming around the impedances.The leads were replaced and the issue was resolved.The manufacturer representative (rep) indicated they would send any further information as they become aware.
 
Manufacturer Narrative
Other relevant device(s) are: product id: 977a260, serial/lot #: (b)(6), ubd: 02-feb-2026, udi#: (b)(4); product id: 97 7a260, serial/lot #: (b)(6); ubd: 07-oct-2025, udi#: (b)(4).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18978987
MDR Text Key338607108
Report Number3004209178-2024-07823
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00763000315467
UDI-Public00763000315467
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/14/2022
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/01/2024
Initial Date FDA Received03/26/2024
Date Device Manufactured07/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H11...."
Patient Outcome(s) Required Intervention;
Patient Age35 YR
Patient SexMale
-
-