Model Number 720185-01
Device Problem Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2024
Event Type  Injury  
Event Description
It was reported that the reservoir of this inflatable penile prosthesis (ipp) migrated from the space anterior of bladder to the midline abdomen.The patient was reported to have a hernia which was not device related; therefore, a revision surgery was performed.All components of the ipp were explanted and replaced.The patient is expected to recover from the revision surgery.No patient complications were reported.
 
Manufacturer Narrative
There was no device available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported patient symptoms are a known risk associated with implant of these devices as indicated in the instructions for use (ifu).Based on the information available, a conclusion code of known inherent risk of device was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18981842
MDR Text Key338658771
Report Number2124215-2024-17922
Device Sequence Number1
Product Code FHW
UDI-Device Identifier00878953005669
UDI-Public00878953005669
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
N970012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/19/2016
Device Model Number720185-01
Device Catalogue Number720185-01
Device Lot Number909854010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/08/2024
Initial Date FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/09/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CX PRECONNECT: UPN: 72404233 / LOT: 913476003
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age75 YR
-
-