ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problems Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
Patient Problem Insufficient Information (4580)
Event Date 02/28/2024
Event Type  Injury  
Event Description
A customer reported a "replace sensor" error message with the adc device and was unable to obtain readings.It was indicated that the customer was provided unspecified treatment by a non-healthcare professional.No further details were provided.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
Sensor (b)(6)has been returned and investigated.Visual inspection has been performed and no issues were observed.Extracted data from returned sensor using approved software.Sensor was found to be in state 5 (indicating normal termination).The sensor plug was properly seated in the mount.Removed the sensor plug and inspected the plug assembly, no failure modes were observed.Sensor was reprogrammed and sim vivo (simulation of the electrical signal produced by the sensor tail) test was performed and all results were within specification.Additional testing has been performed for this issue and poise voltage testing was within specification, indicating the sensor was providing accurate glucose readings.All results were within specification.No malfunction or product deficiency was identified.Therefore, the issue is not confirmed.An extended investigation has been performed for the reported complaint and determined that there was no indication that the product did not meet specification.The dhrs (device history record) for the libre sensor and sensor kit were reviewed, and the dhrs showed the libre sensor and sensor kit passed all tests prior to release.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key18984032
MDR Text Key338657731
Report Number2954323-2024-10058
Device Sequence Number1
Product Code QLG
UDI-Device Identifier05021791001033
UDI-Public(01)05021791001033(17)2025-04-30(10)KTP007877(91)71994-01
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71992-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2024
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-