Model Number 72075-01
Device Problems Device Alarm System (1012); Application Program Problem (2880)
Patient Problem Hyperglycemia (1905)
Event Date 01/01/2024
Event Type  Injury  
Manufacturer Narrative
The most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The customer experienced missing high glucose alarms with freestyle libre 3 application.Adc attempted to replicate the reported issue using similar configuration of samsung galaxy note 10 with android 12 for app version 3.5.1.10363.The reported issue was unable to be replicated and the system functioned as intended.There were no issue identified with the freestyle libre 3 app during replication that would have led to the reported issue.Therefore, issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
An alarm issue was reported with the adc device in use with a samsung galaxy s10, 5g with android operating system version 12.The high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, customer experienced "sugar shock ¿and was unable to self-treat, requiring administration of insulin (dose/type unspecified) provided by their spouse for treatment.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 3 APPLICATION
Type of Device
CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE LTD
range road
witney 12345 -700
UK  12345-7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key18984230
MDR Text Key338658675
Report Number2954323-2024-10060
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
K213996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number72075-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/28/2024
Initial Date FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/28/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-