• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. PATIENT CONNECTOR; DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. PATIENT CONNECTOR; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number 24967
Device Problem Device Difficult to Program or Calibrate (1496)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Manufacturer Narrative
Product event summary: performance data collected from the mobile programmer was received and analyzed.Analysis of the product data /database was able to confirm the customer comment that during implant the implantable pulse generator (ipg) exhibited intermittent telemetry with the mobile programmer and white dots were observed on the electrograms (egm).Excessive significant latency was noted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that during implant the implantable pulse generator (ipg) exhibited intermittent telemetry with the mobile programmer and white dots were observed on the electrograms (egm).The ipg was implanted and remains in use. no patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT CONNECTOR
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18985129
MDR Text Key338663887
Report Number2182208-2024-01315
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00763000060428
UDI-Public00763000060428
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24967
Device Catalogue Number24967
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/20/2024
Initial Date FDA Received03/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
W3SR01 IMPLANTABLE PULSE GENERATOR (IPG)
Patient Age88 YR
Patient SexMale
-
-