The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The date of event is unknown.The date entered in section b3 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
A low readings issue was reported with the abbott diabetes care (adc) device.The customer received lower sensor scan result of "40-45 mg/dl" compared to blood glucose readings obtained on a healthcare professional (hcp) meter of "200-260 mg/dl".It was further reported that the customer was able to self-treat (unspecified and then had contact with an hcp who provided unspecified treatment for the reported issue.No treatment details were provided as the customer refused to troubleshoot further.There was no report of death or permanent impairment associated with this event.A sensor result of 40 mg/dl was compared to a laboratory result of 241 mg/dl and the results, when plotted on a parkes error grid and fell into the "c" zone, showing the difference in values to be clinically significant.It is unknown when these readings were obtained in relation to the reported adverse event.
|