Catalog Number 254501709 |
Device Problems
Peeled/Delaminated (1454); Scratched Material (3020)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/19/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
It was reported that peeling trays and cut block unable to use due to surface damage.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h6 (health effect - clinical code & health effect - impact code).
|
|
Event Description
|
Additional information n received: a.Was there any adverse consequences that affected the patient because of the reported event? no b.Was there a surgical delay? if yes, what is the duration of the delay? no c.Can you please clarify what do you mean by ¿damaged¿? was it worn, cracked, broken into pieces, bent, stripped, cross threaded or any device interaction.Gouge in block.D.Event description states that "cut block unable to use due to surface damage." please clarify what you mean by damaged.Was it broke, cracked, bent or any other deficiency with the device that was not function as intended? deep lacerations.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Search Alerts/Recalls
|
|