• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FARAPULSE, INC. FARADRIVE STEERABLE SHEATH CLEAR; VASCULAR GUIDE-CATHETER SINGLE USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FARAPULSE, INC. FARADRIVE STEERABLE SHEATH CLEAR; VASCULAR GUIDE-CATHETER SINGLE USE Back to Search Results
Lot Number CL12291
Device Problems Leak/Splash (1354); Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Event Description
It was reported that a faradrive steerable sheath clear during insertion for a pulmonary vein isolation de novo procedure, the device was leaking, and blood could be aspirated.The sheath was flushed with a syringe.No other devices introduced, only dilator of the sheath.Leak present on the sheath valve.Air in the syringe during aspiration.The procedure was aborted while the patient was under sedation.The device is expected to be returned.No patient complications reported.
 
Manufacturer Narrative
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FARADRIVE STEERABLE SHEATH CLEAR
Type of Device
VASCULAR GUIDE-CATHETER SINGLE USE
Manufacturer (Section D)
FARAPULSE, INC.
3715 haven avenue
suite 110
menlo park CA 94025
Manufacturer (Section G)
CENTERPOINT SYSTEMS
3338 parkway blvd
west valley city UT 84119
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key18986683
MDR Text Key338708510
Report Number2124215-2024-18267
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeNL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot NumberCL12291
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/18/2024
Initial Date FDA Received03/27/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/28/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-