Model Number 71940-01
Device Problems Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
Patient Problems Hyperglycemia (1905); Loss of consciousness (2418)
Event Date 02/23/2024
Event Type  Injury  
Event Description
An error message was reported with the abbott diabetes care (adc) device.The customer received "replace sensor" message and was unable to obtain glucose readings.As a result, the customer experienced loss of consciousness.The customer was unable to self-treat due to symptoms requiring hcp administration of insulin.No further treatment was reported.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
Sensor (b)(6) has been returned and investigated.The sensor plug is properly seated and no physical damage was observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.Perform sim vivo testing (simulation of the electrical signal produced by the sensor tail) and poise voltage testing.All results were within specification.Therefore, issue is not confirmed.No malfunction or product deficiency was identified.Dhrs (device history review) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 14 DAY
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key18987316
MDR Text Key338720415
Report Number2954323-2024-10201
Device Sequence Number1
Product Code PZE
UDI-Device Identifier00357599001018
UDI-Public(01)00357599001018(17)240630(10)231012P(91)71940-01
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71940-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2024
Initial Date Manufacturer Received 03/05/2024
Initial Date FDA Received03/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
Patient SexMale
-
-