Model Number 6300
Device Problem Inaccurate Flow Rate (1249)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2024
Event Type  malfunction  
Manufacturer Narrative
One device was returned for evaluation.Visual inspection revealed no physical damage.Event history log confirmed no miss setting or other errors related to delivery failures were identified.Functional testing could not replicate the reported issue; the pump accuracy was within specification.The root cause could not be determined as there was no problem with the pump accuracy.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.
 
Event Description
It was reported that the device flow rate accuracy error was more than 6 percent.The issue was discovered during inspection.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-LEGACY PCA AMBULATORY INFUSION PUMP
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18987737
MDR Text Key338725045
Report Number3012307300-2024-01717
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6300
Device Catalogue Number21-6300-09JP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2024
Date Manufacturer Received02/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-